Australian Standard Gas Outlet-DMG11-8

Product Introduction:

The medical gas terminal is a standardized quick-connect interface installed at points of use such as patient rooms, ICUs, and operating theaters, connecting equipment like ventilators and suction units to the central gas supply. Non-interchangeable configurations prevent misconnection between different gases (oxygen, vacuum, compressed air). An integrated self-sealing mechanism automatically shuts off gas flow upon disconnection, eliminating leaks and ensuring patient and staff safety.


TECHNICAL SPECIFICATIONS

1. Gas Type:O₂,MEDICAL AIR,VAC,N₂O,N₂,CO₂are available

2. Available in wall type with box,console type on bed head units and hose barb type

3. 100% pressure and leakage tested

4. Quick connections with no risk of gas cross-connection

5. Color-coded according to ISO 32

6. Suitable ambient temperature:-20°C to 60°C,ambient humidity:10% to 95%

7. The working pressure for oxygen,air(breathing),nitrous oxide,and carbon dioxide is 0.4 MPa

8. The working pressure for nitrogen and instrument air is 0.8 MPa

9. The negative pressure working pressure is -0.075 MPa

10. The maximum gas supply flow rate is more than 100L/min;the maximum suction flow rate is more than 40L/min


INQUIRE IMMEDIATELY